BMB-II-"Laminar-S."-0,9 SAVVY SL Class II

 

Article 1Е-B.008-09.0

The microbiological safety cabinet Class II SAVVYSL is designed for work with agents of biosafety levels I, II and III. It provides product, operator and environment protection.

SavvySLis a new generation of the microbiological safety cabinets that inherited all the advantages of the previous models.

KEY BENEFITS

POWER DRIVEN FRONT SASH. 

Five specified positions of the front sash:

  • Operating position is settled  in accordance with EN 12469
  • Completely closed position is used in Stand-by Mode
  • Down position – for cleaning and disinfection
  • Up position – for loading the working chamber
  • Stop before closing 

FRONT SASH STOP PRIOR TO CLOSING. Automatic stop of the front sash at 40 mm height from the surface prior to complete closing of the front opening prevents any injury of the operator’s hands.

STOP AHEAD OF OBSTACLE. To ensure maximum safety, the moving front sash automatically stops in case of contact with an object or a hand of the operator.

MICROPROCESSOR CONTROL SYSTEM.  The cabinet is featured with a microprocessor control system that immediately informs an operator on the reduction of the level of the cabinet protection via alarm messages displayed on the touchscreen and via an audible and visible alert.

The fan motor control system SintelL-1 minimizes the power consumption of the operating cabinet as well as reduces the level of acoustic and electromagnetic noise. 

The system of static airflow stabilization AIS LS automatically maintains the air balance in the work chamber increasing the fan rpm as filters get clogged. Upon reaching the clogging threshold, the alarm is activated. 

Mode monitoring system featured with audible and visible alarm indicates any disturbance of the air balance in the work chamber.

AUDIBLE AND VISIBLE ALARM SYSTEM.  Alarms are automatically activated in case the airflow parameters deviate from the preset values.

LOW LEVEL OF NOISE.

TOUCHSCREEN CONTROL PANEL.  It ensures clear visualization of operating modes and simple control of the cabinet as well as provides the user with more service and maintenance information. The screen allows for work in gloves and disinfection with liquids, i.e. hydrogen peroxide.

 

      

 

MAIN CHARACTERISTICS

Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1

- for particles of 0.5µm and more

- for particles of 5.0µm and more

 

ISO 5 

ISO M (20; ≥5µm); LSAPC 

Cabinet class according to ЕN 12469

II

Class of the installed НЕРА-filters according to EN 1822-1

H14

Average velocity of the inflow though the work opening, m/s

0,47±0,03

Average downflow velocity in the working chamber, m/s 

0,35+0,01

Clean air inflow volume, m3/h

656-674

Air outflow volume, m3/h

289-328

Air recirculation rate in the cabinet, %

≈70/30

 

MAIN PARAMETERS AND DIMENSIONS

Article

1Е-B.008-09.0

Dimensions of the cabinet with the stand in assembly (WxDxH), mm

 1000х810х2095

Dimensions of the cabinet with stand and completely opened front sash (WхDхH).mm 1000х810х2180

Dimensions of the working chamber (WxDxH), mm

905х610х700

Weight of the cabinet with the stand in assembly (net), kg, not more than

215

Cabinet's power supply parameters:

- voltage, V~

- frequency, Hz

 

220-240

50

Maximum power consumption (without the built-in outlets load), W, not more than

680/220*

Total acceptable load on the built-in outlets, W, not more than

1000

* The power consumption was measured in the cabinet with newly installed (uncontaminated) HEPA filters.

ADDITIONAL OPTIONS

  • ULPA filters 
  • Exhaust hood for connecting the cabinet to the external exhaust system 
  • Adjustable stand 
  • Additional electric sockets 
  • Technical gas tap with electromagnetic valve*
  • Inflammable gas tap with electromagnetic valve*
  • Vacuum tap with electromagnetic valve*

* The electromagnetic valve automatically locks gas and vacuum supply in case of power cutoff or cabinet switching-off.

DIMENSION DRAWING

 