Biological Safety Cabinet BMB-II-"Laminar-S."-1,8 NEOTERIC B2 Class II

Article 2E-B.003-18

Class II Type B2 Biosafety Cabinet ensures environment, product, and personal protection during lab procedures where the circulation of filtered air is restricted due to the existence of possible chemical fumes that are released during the process. Since the filter does not catch these chemical fumes, it can be of potential danger to the operator if they are re-circulated in the workspace. Therefore, Class II Type B2 Biosafety Cabinets are ducted to a blower that is located on the outside of the cabinet, for total exhaust.

 

 ATTENTION! The cabinet must be connected to an individual active exhaust system with a flow rate of 1135-1205 m3/h. The room where the cabinet is installed must have a supply ventilation system with a flow rate of not less than 1000 m3/h.

Version for print

Microbiological safety cabinet BMB-II-“Laminar-S.”-1,8 NEOTERIC B2

Class II



Article 2Е-В.003-18
Dimensions of the assembled cabinet without exhaust hood, mm (WхDхH)* 1800х810х2009
Dimensions of the assembled cabinet with exhaust hood, mm (WхDхH)* 1800х810х2220
Dimensions of the cabinet’s work chamber, mm (WхDхH) 1705х610х705
Mass of assembled cabinet with the stand and , kg, not more than 370
Input power of the cabinet (exclusively of the load on the built-in outlets), W, not more than 960/320**
Maximum load allowed on the sockets, W 2000
Air volume supplied to the work chamber of the cabinet, m3/h 1070-1100
Air volume exhausted from the cabinet, m3/h 1590-1690

* dimensions do not account for outstanding supports

**The power consumption was measured in the cabinet with newly installed (uncontaminated) HEPA filters.


MAIN CHARACTERISTICS


Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1
-for particles of 0.5µm and more ISO 5
-for particles of 5.0µm and more ISO M (20; ≥5µm); LSAPC
Class of the cabinet according  to EN 12469-2010, NSF/ANSI 49 II
Type of the cabinet according to NSF/ANSI 49 B2
Class of the installed in the cabinet НЕРА-filters according to EN 1822-1 Н14
Class of the installed preliminary filter according to EN 779 G4
Average downflow velocity in the work chamber, m/s 0,35+0,01
Average velocity of the air inflow though the work opening, m/s, not less than 0,47±0,03
Illuminance level in the work zone, lux, not less than 2000
Air recirculation rate in the cabinet, % missing
Dimensional drawing
Work Chamber
UV Unit
Supply and Exhaust Air Filtration System
Control System Components
Additional Components
Components for Testing the supply HEPA Filter Integrity
Components for Testing the exhaust HEPA Filter Integrity