Biological Safety Cabinet BMB-II-"Laminar-S."-1,2 SAVVY B2 Class II

Article 2E-B.003-12

Class II Type B2 Biosafety Cabinet ensures environment, product, and personal protection during lab procedures where the circulation of filtered air is restricted due to the existence of possible chemical fumes that are released during the process. Since the filter does not catch these chemical fumes, it can be of potential danger to the operator if they are re-circulated in the workspace. Therefore, Class II Type B2 Biosafety Cabinets are ducted to a blower that is located on the outside of the cabinet, for total exhaust.

Version for print

Microbiological safety cabinet BMB-II-“Laminar-S.”-1,2 SAVVY B2

Class II


Article 2Е-В.003-12
Dimensions of the assembled cabinet with exhaust hood, mm (WхDхH)* 1200х810х2325
Dimensions of the cabinet’s work chamber, mm (WхDхH) 1105x610x705
Mass of assembled cabinet with the stand, kg, not more than 270
Input power of the cabinet (exclusively of the load on the built-in outlets), W, not more than 730
Maximum load allowed on the sockets, W 1000
Air volume supplied to the work chamber of the cabinet, m3/h 795-820
Air volume exhausted from the cabinet, m3/h 1180-1220

* dimensions do not account for outstanding supports


ATTENTION! The cabinet must be connected to an individual active exhaust system with a flow rate of 1135-1205 m3/h. The room where the cabinet is installed must have a supply ventilation system with a flow rate of not less than 1000 m3/h.


MAIN CHARACTERISTICS


Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1
-for particles of 0.5µm and more ISO 5
-for particles of 5.0µm and more ИСО М (20; ≥ 5 мкм);LSAPC
Class of the cabinet according  to EN 12469-2010, NSF/ANSI 49 II
Type of the cabinet according to NSF/ANSI 49 B2
Class of the installed in the cabinet НЕРА-filters according to EN 1822-1 H14
Class of the installed preliminary filter according to EN 779 G4
Average downflow velocity in the work chamber, m/s 0,35+0,01
Average velocity of the air inflow though the work opening, m/s, not less than 0,53±0,02
Illuminance level in the work zone, lux, not less than 1000
Air recirculation rate in the cabinet, % missing

* dimensions do not account for outstanding supports

Dimensional drawing
WorkChamber
UV unit
Supply and Exhaust Air Filtration System
Control System Components
Additional Components
Components for Testing the supply HEPA Filter Integrity
Components for Testing the exhaust HEPA Filter Integrity