Biological Safety Cabinet BMB-II-“Laminar-S”-1,5

Article 1Е-В.003-15.0

Air flow scheme - Microbiological Safety Cabinet Class 2 (Biological Safety Cabinet Class 2)

 

Class II Type B2 Biosafety Cabinet ensures environment, product, and personal protection during lab procedures where the circulation of filtered air is restricted due to the existence of possible chemical fumes that are released during the process. Since the filter does not catch these chemical fumes, it can be of potential danger to the operator if they are re-circulated in the workspace. Therefore, Class II Type B2 Biosafety Cabinets are ducted to a blower that is located on the outside of the cabinet, for total exhaust.

ATTENTION! The cabinet must be connected to an individual active exhaust system with a flow rate of 1434-1520 m3/h. The room where the cabinet is installed must have a supply ventilation system with a flow rate of not less than 1000 m3/h.

Main  characteristics of  the  biological   safety   cabinet   classII  type  B2  BMB-II-“Laminar-S”-1,5

Article

1Е-В.003-12.0

Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1

-for particles of 0.5µm and more

-for particles of 5.0µm and more

 

ISO 5 

ISO M (20; ≥5µm); LSAPC

Class of the cabinet according  to EN 12469-2010, NSF/ANSI 49 II
Type of the cabinet according to NSF/ANSI 49 B2
Class of the installed in the cabinet НЕРА-filters according to EN 1822-1 Н14
Class of the installed preliminary filter according to EN 779 G4
Average downflow velocity in the work chamber, m/s 0,35+0,01
Average velocity of the air inflow though the work opening, m/s, not less than 0,47±0,03
Illuminance level in the work zone, lux, not less than 2000
Air recirculation rate in the cabinet, % missing

Main parameters and dimensions of the biological safety cabinet class II type B2 BMB-II-“Laminar-S”-1,5

Dimensions of the assembled cabinet, mm (WхDхH) 1500х810х2220*
Dimensions of the cabinet’s work chamber, mm (WхDхH) 1405х610х705
Input power of the cabinet (exclusively of the load on the built-in outlets), W, not more than 730
Maximum load allowed on the sockets, W 1000
Air volume supplied to the work chamber of the cabinet, m3/h 1008-1036
Air volume exhausted from the cabinet, m3/h 1434-1520

 * - Reference dimensions that may slightly change in the process of manufacturing

 

 

Additional Options to biological safety cabinet class II type B2

  • Multi-purpose taps;
  • Extra outlets;
  • LED lighting of the working chamber.

Dimensional drawing of the BMB-II-“Laminar-S”-1,5



 

 