BMB-II-Laminar-S-1,8 SAVVY

Article 1Е-В.002-18.0

Microbiological safety cabinet class II SAVVY (code 222.180)

 Download brochure

 

The biological safety cabinet Class II SAVVY is designed for work with agents of biosafety levels I, II and III. It provides product, operator and environment protection.

Savvy is a new generation of the microbiological safety cabinets that inherited all the advantages of the previous model line Neoteric.  A set of constructive and layout solutions have drastically improved exploitational characteristics of the new cabinet. 

KEY BENEFITS

LOW POWER CONSUMPTION 0.10 kW 

ENERGY SAVING EC FANS

LOW HEAT EMMISSION 30 W

EASY AND CONVENIENT AIRFLOW VELOCITY SETTING

The system provides separate inflow and downflow control as well as automatic air balance control. No need of mechanical air balance setting significantly decreases maintenance time when validating, changing filters and performing periodic
verifications.

HIGH MAINTENANCE ACCURACY OF PRESET AIRFLOW VELOCITY with any level of filters clogging or changing ambient conditions (humidity, temperature, pressure)

THE USE OF TOUCHSCREEN made displaying modes more demonstrative, the cabinet control easier and provided the user with more service information.

LARGE WORKING CHAMBER WITH MODEST TOTAL DIMENSIONS 

EASY AND SAFE FILTER CHANGING thanks to the layout solution and a new pressure system minimizing the filter damage risk during their installation.

BROAD BASIC CONFIGURATION

  • unique UV-unit,
  • visual-audio REFLEXLS alarm,
  • optical sensors of the UV-unit and front glass position,
  • UV-unit and front
  • glass dampers,
  • outlets unit in the working chamber,
  • working chamber LED lighting,
  • removable armrest, frame with a footrest.
  • UV light is located in the pull-out unit outside the work chamber and does not interfere with the airflow at work.
 
 MAIN CHARACTERISTICS

Air cleanliness class in the working chamber of the cabinet in terms of concentration of airborne particles (aerosols) according to ISO 14644-1

 5 ISO

Cabinet class according to ЕN 12469, NSF/ANSI 49

II

Cabinet type according to NSF/ANSI 49

A2

Class of the installed НЕРА-filters according to EN 1822-1

H14

Average velocity of the inflow though the work opening, m/s

0,47±0,03

Average downflow velocity in the working chamber, m/s 

0,35+0,01

Clean air inflow volume, m /h

  1210-1245   

Air outflow volume, m3/h

510-580

Air recirculation rate in the cabinet, %

≈70

Illuminance level in the working zone, lux, not less than

2000

Noise level at 1m distance from the cabinet, dB, not more than 

55*

 *in operating mode (fans and lighting turned on) in testing laboratory environment.

MAIN PARAMETERS AND DIMENSIONS

Article

1Е-В.002-18.0

Dimensions of the cabinet with the stand in assembly (WxDxH), mm

 1800x770x2095

Dimensions of the working chamber (WxDxH), mm

1705x610x750

Cabinet's power supply parameters:

-voltage, V~

-frequency, Hz

 

220-240

50

Maximum power consumption (without the built-in outlets load), W, not more than

770/220**

Total acceptable load on the built-in outlets, W, not more than

1000

Fan number, pieces

 4

** The power consumption was measured in the cabinet with newly installed (uncontaminated) HEPA filters.

ADDITIONAL OPTIONS

  • Multi-purpose taps
  • Extra outlets in the working chamber
  • ULPA-filters set (supply and exhaust)
  • Hood to connect the cabinet to external exhaust system

Dimensional drawing of SAVVY ( code 222.180)