Biological Safety Cabinet BMB-II-“Laminar-S”-1,8

Article 1Е-В.003-18.0

 

Class II Type B2 Biosafety Cabinet ensures environment, product, and personal protection during lab procedures where the circulation of filtered air is restricted due to the existence of possible chemical fumes that are released during the process. Since the filter does not catch these chemical fumes, it can be of potential danger to the operator if they are re-circulated in the workspace. Therefore, Class II Type B2 Biosafety Cabinets are ducted to a blower that is located on the outside of the cabinet, for total exhaust.

 

 ATTENTION! The cabinet must be connected to an individual active exhaust system with a flow rate of 1135-1205 m3/h. The room where the cabinet is installed must have a supply ventilation system with a flow rate of not less than 1000 m3/h.

Main  characteristics of  the  biological   safety   cabinet   classII  type  B2  BMB-II-“Laminar-S”-1,8 

Article

1Е-В.003-18.0

Air cleanliness class in the working chamber of the cabinet for suspended particles content according to ISO 14644-1


5 ISO
Class of the cabinet according  to EN 12469-2010, NSF/ANSI 49 II
Type of the cabinet according to NSF/ANSI 49 B2
Class of the installed in the cabinet НЕРА-filters according to EN 1822-1 Н14
Class of the installed preliminary filter according to EN 779 G4
Average downflow velocity in the work chamber, m/s 0,35+0,01
Average velocity of the air inflow though the work opening, m/s, not less than 0,47±0,03
Illuminance level in the work zone, lux, not less than 2000
Air recirculation rate in the cabinet, % missing

Main parameters and dimensions of the biological safety cabinet class II type B2 BMB-II-“Laminar-S”-1,8

Dimensions of the assembled cabinet, mm (WхDхH) 1800х810х2325
Dimensions of the cabinet’s work chamber, mm (WхDхH) 1705х610х705
Mass of assembled cabinet with the stand and , kg, not more than 370
Input power of the cabinet (exclusively of the load on the built-in outlets), W, not more than 960
Maximum load allowed on the sockets, W 1000
Air volume supplied to the work chamber of the cabinet, m3/h 1070-1100
Air volume exhausted from the cabinet, m3/h 1590-1690

*The power consumption was measured in the cabinet with newly installed (uncontaminated) HEPA filters.

Additional Options to biological safety cabinet class II type B2

  • Multi-purpose taps
  • Extra outleta
  • LED lighting of the working chamber

Dimensional drawing of the BMB-II-“Laminar-S”-1,8