BMB-II-Laminar-S-1,5 NEOTERIC

Article 1Е-В.001-15.0

    

Download brochure

 

 

Distinctive features  

 

 

Video

 

 

MAIN CHARACTERISTICS

Air cleanliness class in the working chamber of the cabinet in terms of concentration of airborne particles (aerosols) according to ISO 14644-1, not less than

 5 ISO

Cabinet class according to ЕN 12469, NSF/ANSI 49

II

Cabinet type according to NSF/ANSI 49

A2

Class of the installed НЕРА-filters according to EN 1822-1

H14

Average velocity of the inflow though the work opening, m/s

0,47±0,03

Average downflow velocity in the working chamber, m/s

0,35+0,01

Illuminance level in the working zone, lux, not less than

    1000    

Air recirculation rate in the cabinet, %

≈70

 MAIN PARAMETERS AND DIMENSIONS 

Article

1Е-В.001-15.0

Dimensions of the cabinet with the stand in assembly (WxDxH), mm

 1500x770x2150

Dimensions of the working chamber (WxDxH), mm

1405x610x750

Weight of the cabinet with the stand in assembly (net), kg, not more than

300

Power consumption (without the built-in outlets load), W, not more than

330/160*

Total acceptable load on the built-in outlets, W, not more than

1000

Air volume supplied to the working chamber,  m3 /h

1008-1036

      *The power consumption with newly installed uncontaminated HEPA filters.
  • The air flow in the working chamber – vertical downward unidirectional (laminar).
  • The air flow in the front opening – directed into the front table grille, creating an air curtain.
  • There is no risk of plenum contamination (the internal high pressure plenum is completely surrounded by low pressure zones)

ADDITIONAL OPTIONS

  • multi-purpose taps
  • extra outlets
  • ULPA-filters
  • LED lighting of the working chamber
  • Fume hood to connect the cabinet to external exhaust system

DIMENSION DRAWING

     

Class II Type A2 Biological Safety Cabinets 

MODELS 

BMB-II-Laminar-S-0,9

BMB-II-Laminar-S-1,2

BMB-II-Laminar-S-1,8

BMB-II-Laminar-S-1,2 SAVVY